The Medical Devices Regulation sets requirements according to device class. More specifically, medical devices are defined as class I, IIa, IIb, or III. In turn, this determines certain requirements, and which conformity assessment procedure the manufacturer must follow.
In this guide, we present the different classes for medical devices, explain how medical devices must be classified, delve into the classification rules, and explore the implications of each classification.
Content Overview
FREE CONSULTATION CALL (US, EU & UK)
- Request a free 30-minute call with Ivan Malloci to learn how we can help you with:
- Find product requirements
- Certification and labeling
- Lab testing
Classification rules for medical devices
The Medical Devices Regulation requires medical devices to be classified into one of the four classes:
- Class I devices
- Class IIa devices
- Class IIb devices
- Class III devices
This is necessary to confirm the conformity assessment procedure which needs to be undertaken as not all classes share the same conformity assessment procedure. Additionally, the class of the device affects the labelling information accompanying the product.
The conformity assessment procedure explains the route that manufacturers should take to demonstrate that their product conforms to the requirements of the Medical Devices Regulation. Note that some conformity assessment procedures specify the involvement of a notified body.
A medical device’s class can be determined if it satisfies the relevant rules contained in Annex VIII of the Medical Devices Regulation. Below we provide some examples of the classification rules taken from the regulation, which classes they are related to, and their description.
Rule | Class | Description | |||
Rule 1 | Class I | This rule states that non-invasive devices are classed as Class I devices unless another rule states otherwise. | |||
Rule 12 | Class IIa, Class IIb | This rule states that active devices used to administer or remove substances (e.g. medicinal products) to or from the body are classed as Class IIa devices.
However, it would be classified as Class IIb if the active device functions in a potentially hazardous manner. |
|||
Rule 19 | Class III | This rule states that devices using nanomaterials that present a high or medium potential for internal exposure are classified as Class III devices. |
The European Commission published a guide on the classification rules, which includes examples of products related to each rule. To view the guide, enter into the European Commission’s site containing guidance documents concerning the Medical Devices Regulation and search for the document entitled “Guidance on classification of medical devices”.
The section entitled “4.2 General explanation of rules/practical issues/examples” in the guide contains an explanation of each classification rule.
Class I medical device
This section highlights the classification rules related to Class I medical devices, provides product examples, and covers the conformity assessment procedure required.
Classification rules
The following rules are relevant to a Class I determination for a medical device:
- Non-invasive devices (rules 1, 2, and 4)
- Invasive devices (rules 5 and 6)
- Active devices (rules 10, 11, and 13)
- Special rules (rule 18)
Product examples
The following are examples of Class I medical devices and the applicable classification rule:
- Stethoscopes (Rule 1)
- Corrective spectacle frames (i.e. glasses) and lenses in frames (Rule 1)
- Cups and spoons specifically intended for administering medicines (Rule 2)
- Sticking plasters (Rule 4)
- Examination gloves (Rule 5)
- Scalpels and scalpel handles (Rule 6)
Requirements
The following Class I devices should adhere to the conformity assessment procedure in Annex IX or Annex XI if they:
- Are sold in a sterile condition
- Have a measuring function
- Are reusable surgical instruments
For all other class I devices, the regulation only specifies the issuance of the Declaration of Conformity and technical documentation.
Summary
Requirement | Notes | |||
Notified body’s identification number label | Class I devices (sterilized devices, devices with measuring functions, or reusable surgical instruments)
The notified body identification number needs to be displayed along with the CE mark Other Class I devices Not required |
|||
Documentation | Class I devices (sterilized devices, devices with measuring functions, or reusable surgical instruments)
The required documents vary depending on the conformity assessment procedure selected. Examples of required documentation include: a. Declaration of conformity b. Technical documentation c. Test reports Other Class I devices a. Declaration of Conformity b. Technical documentation |
|||
Notified body | Class I devices (sterilized devices, devices with measuring functions, or reusable surgical instruments)
Required Other Class I devices Not required |
|||
Other requirements unique to Class I devices | Preparation of a post-market surveillance report is required |
Class IIa medical devices
This section highlights the classification rules related to Class IIa medical devices, provides product examples, and covers the conformity assessment procedure required.
Classification rules
The following rules are relevant to a Class IIa determination for a medical device:
- Non-invasive devices (rules 2 – 4)
- Invasive devices (rules 5 – 8)
- Active devices (rules 9 – 12)
- Special rules (rules 16 – 17 and 19 – 21)
Product examples
The following are examples of Class IIa medical devices and the applicable classification rule:
- Syringes for infusion pumps (Rule 2)
- Short term corrective contact lenses (Rule 5)
- External hearing aids (Rule 9)
- Digital x-ray detectors for recording images (Rule 17)
- Dental filling materials (Rule 19)
Requirements
Class IIa devices should adhere to the conformity assessment procedure specified in either:
a. Annex IX
b. Annex XI
Summary
Requirement | Notes | |||
Notified body’s identification number label | The notified body identification number needs to be displayed along with the CE mark | |||
Documentation | Annex IX
There are numerous types of documentation specified in the Annex. Examples of required documentation include: a. Declaration of conformity b. Technical documentation c. Test reports d. Documentation on the manufacturer’s post-market surveillance system e. Documentation on the clinical evaluation plan Annex XI There are numerous types of documentation specified in the Annex. Examples of required documentation include: a. EU type-examination certificate b. Technical documentation c. EU declaration of conformity |
|||
Notified body | Required | |||
Other requirements unique to Class IIa devices | 1. A summary of safety and clinical performance to be drawn up for implantable devices is required
2. A periodic safety update report is required |
Class IIb medical devices
This section highlights the classification rules related to Class IIb medical devices, provides product examples, and covers the conformity assessment procedure(s) required.
Classification rules
The following rules are relevant to a Class IIb determination for a medical device:
- Non-invasive devices (rules 2 – 4)
- Invasive devices (rules 5 – 8)
- Active devices (rules 9 – 12)
- Special rules (rules 15 – 16 and 19 – 21)
Product examples
The following are examples of Class IIb medical devices and the applicable classification rule:
- Dressings intended for burns having breached the dermis (Rule 4)
- Long term corrective contact lenses (Rule 5)
- Lung ventilators (Rule 9)
- Condoms and femidoms (Rule 15)
- Eye drops for hydration (Rule 21)
Requirements
Class IIb devices should adhere to the conformity assessment procedure specified in either:
a. Annex IX
b. Annex X and Annex XI
Summary
Requirement | Notes | |||
Notified body’s identification number label | The notified body identification number needs to be displayed along with the CE mark | |||
Documentation | Annex IX
There are numerous types of documentation specified in the Annex. Examples of required documentation include: a. Declaration of conformity b. Technical documentation c. Test reports d. Documentation on the manufacturer’s post-market surveillance system e. Documentation on the clinical evaluation plan Annex X and Annex XI There are numerous types of documentation specified in the Annex. Examples of required documentation include: a. EU type-examination certificate b. Technical documentation c. EU declaration of conformity d. Scientific opinions and reports e. Test reports |
|||
Notified body | Required | |||
Other requirements unique to Class IIb devices | 1. A summary of safety and clinical performance to be drawn up for implantable devices is required
2. A periodic safety update report is required |
Class III medical devices
This section highlights the classification rules related to Class III medical devices, provides product examples, and covers the conformity assessment procedure(s) required.
Classification rules
The following rules are relevant to a Class III determination for a medical device:
- Non-invasive devices (rule 3)
- Invasive devices (rules 6 – 8)
- Active devices (rules 9 and 11)
- Special rules (rules 14, 15, 18, 19, 21, and 22)
Product examples
The following are examples of Class III medical devices and the applicable classification rule:
- IVF cell media without human albumin (Rule 3)
- Cardiovascular catheters (Rule 6)
- Remote monitoring devices for active implantable devices (Rule 9)
- Bone cement with antibiotics (Rule 14)
Requirements
Class III devices should adhere to the conformity assessment procedure specified in either:
a. Annex IX
b. Annex X and Annex XI
Summary
Requirement | Notes | |||
Notified body’s identification number label | The notified body identification number needs to be displayed along with the CE mark | |||
Documentation | Annex IX
There are numerous types of documentation specified in the Annex. Examples of required documentation include: a. Declaration of conformity b. Technical documentation c. Test reports d. Documentation on the manufacturer’s post-market surveillance system e. Documentation on the clinical evaluation plan Annex X and Annex XI There are numerous types of documentation specified in the Annex. Examples of required documentation include: a. EU type-examination certificate b. Technical documentation c. EU declaration of conformity d. Scientific opinions and reports e. Test reports |
|||
Notified body | Required | |||
Other requirements unique to Class III devices | 1. There are specific recordkeeping requirements concerning the unique device identification (UDI) of Class III implantable devices
2. A summary of safety and clinical performance is required 3. A periodic safety update report is required |